Aligning The Centrifuge; Mains Connection - Thermo Scientific SL40 Instruction Manual

Hide thumbs Also See for SL40:
Table of Contents

Advertisement

• The supporting structure must be stable and free of resonance, for example a level laboratory
bench.
• The supporting structure must be suitable for horizontal setup of the centrifuge.
• The centrifuge should not be exposed to heat and strong sunlight.
• The set-up location must be well-ventilated at all times.

Aligning the Centrifuge

The alignment of the centrifuge must be checked every time after moving it to a different location.
The supporting structure must be suitable for horizontal setup of the centrifuge.

Mains Connection

Figure 2-2. Mains connections
1. Turn off the power supply switch on the back (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
Thermo Scientific
WARNING UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
CAUTION If the centrifuge isn't level, imbalances can occur and the centrifuge can be
damaged.
Do not place anything under the feet to level the centrifuge.
Mains plug
Main Power Switch
2
Before use
Aligning the Centrifuge
SL40 / 40R
2-3