Feature
Environmental conditions
permissible ambient temperature
Overvoltage category
Pollution degree
Heat dissipation
IP
running time
maximum speed n
max
minimum speed n
max
maximum RCF value at n
max
maximum kinetic energy
noise level at maximum speed
temperature setting range
Dimensions
Height
Width
Depth
Table top height
weight without rotor
Thermo Scientific
Value
- use in interior spaces
- altitudes of up to 2,000 m above sea level
- max. relative humidity 80% up to 31 °C; decreasing linearly up to 50%
relative humidity at 40 °C.
+2° C to +35° C
II
2
ventilated
230V
120V
5800 BTU/h
4776 BTU/h
20
unlimited
15200 rpm (depending on the rotor)
300 rpm
25314 xg
< 62,5 kJ
< 64dB (A)
ventilated
360 mm
550 mm
670 mm
310 mm
86 kg
1
Introduction and Description
Technical Data
+2° C to +35° C
II
2
refrigerated
230V
120V
6653 BTU/h
4776 BTU/h
20
unlimited
15200 rpm (depending on the rotor)
300 rpm
25314 xg
< 62,5 kJ
< 64dB (A)
-10 °C to +40 °C
refrigerated
360 mm
745 mm
670 mm
310 mm
116 kg