Aligning the Centrifuge

The horizontal alignment of the centrifuge must be checked every time after moving it to a different
location.
The supporting structure must be suitable for horizontal setup of the centrifuge.

Mains Connection

Figure 2-2. Mains Connection
1. Turn off the power supply switch on the back (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
3. Check whether the cable complies with the safety standards of your country.
4. Make sure that the voltage and frequency correspond to the figures on the rating plate.
5. Establish the connection to the power supply with the connecting cable.
Thermo Scientific
WARNING UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
CAUTION If the centrifuge isn't level, imbalances can occur and the centrifuge can be
damaged.
Do not place anything under the feet to level the centrifuge.
Main Power Switch
Mains plug
2
Before use
Aligning the Centrifuge
SL 16R
2-3