U5310A Front Panel Features
Front Panel Connectors
Please note that analog Input IN 1 and IN 2 have a maximum voltage limitation of
±1.5 Vmax while using the default 1V full scale range (±0.5 Vmax if using 0.25V
FSR).
Connector Type
TRG IN
IN 1, 2
JTAG
TRG OUT
I/O 1, 2, 3
CLK IN
REF IN
Digitizer can usually work with signals present at the external reference and
clock inputs (REF IN and CLK IN). However, to ensure the best performance, or if
the calibration is found to be unreliable, it is recommended to remove such
signals when working with internal clock.
Keysight U5310A Startup Guide
Description
MMCX
External trigger input, 50 Ω DC terminated, ±5 V range.
female
SMA
Analog signal inputs, DC-coupled and 50 Ω terminated.
female
The input full scale ranges are selectable:
Range
0.25 V
1 V
Micro
Provides connection to the DPU for use with the U5340A
USB
FPGA Development Kit. (Not supported)
MMCX
Trigger Out signal. User selectable from several functions.
female
MMCX
User configurable Input / Output signal. 3.3 V CMOS and
female
TTL compatible.
MMCX
External clock input. AC coupled and 50 Ω terminated, sig-
female
nal level: +5 to +15 dBm. (Please refer to product user
manual or
datasheet
MMCX
External reference clock input, AC coupled and 50 Ω ter-
female
minated. It can accept a 100 MHz signal from -3 to +3 dBm.
Maximum Input
±0.5 Vmax
±1.5 Vmax
for supported frequency values).