Usage
EN-
Protect the unit from contamination when
❯
transporting it from the treatment chair to the
US
reprocessing location.
11.5
Manual cleaning, disinfection
and drying
NOTICE
Damage to the device due to incorrect
cleaning and disinfection
Only clean the surface of the unit.
❯
Only use disinfection and cleaning
❯
agents specified or approved by Air
Techniques and the EPA.
Use combined cleaning agents and
❯
disinfectants which do not contain
chlorine, solvents, strong bases (pH
>11), or oxidizing agents.
Do not use any aggressive or abrasive
❯
cleaning agents.
Only clean the unit using wipe disinfec-
❯
tion.
Do not clean the unit by submerging or
❯
spraying in combination with disinfec-
tant.
Do not subject the unit to steam sterili-
❯
zation.
The outer surfaces of the applied part can be
cleaned and disinfected manually using cleaning
and disinfection wipes.
Cleaning
Thoroughly wipe down the outer surfaces for 1
❯
minute with a cleaning wipe.
Then allow for 1 minute exposure to the agent.
Check to make sure that no soiling is visible
❯
any longer.
Disinfecting
Thoroughly wipe down the outer surfaces for
❯
1 minute with a disinfection wipe.
Repeat this step with a new disinfection wipe
❯
for 2 minutes.
This means that the entire disinfection step is
performed for 3 minutes.
Final rinse
After the action time prescribed by the manufac-
turer has expired:
Rinse all components under running water for
❯
at least 30 seconds (temperature
< 35 °C / < 95 °F).
28
Drying
Allow the device to air-dry.
❯
The device must be completely dry before a
new hygienic protective cover is pulled on.
2121100020L29 2101V007