Installation
EN-
Testing the device
US
An X-ray image can be recorded to test whether
the unit has been connected correctly.
Select the Test tab.
❯
Select the unit from the Registered Units list.
❯
Select the mode.
❯
Click on Scan Image.
❯
Record an X-ray image.
❯
8.3
Commissioning tests
The required tests (e.g. acceptance test) must be
done in accordance with local rules and regula-
tions.
Find out which tests are required.
❯
Carry out testing in accordance with local rules
❯
and regulations.
Check image quality
We recommend repeating the test every 3
months. In the event that this requirement devi-
ates from national requirements, please follow
and apply the corresponding national require-
ments.
22
The image quality test is carried out with
VisionX Inspect (see VisionX handbook, order
number: E7303 (2110200001).
The following aids are required:
– VisionX Inspect imaging software
– Test phantom for 2D X-rays (see "3.3 Optional
items").
Positioning the test phantom
On the right-hand side of the insertion opening
❯
push the sensor as far as it will go into the test
phantom. The active side of the sensor must
be facing toward the X-ray emitter.
In addition, a small amount of cotton wool
wadding can be inserted on the side
between the sensor and the test phantom
to secure the sensor in position.
2121100020L29 2101V007