TROUBLESHOOTING

Error/
Error Display
The display is empty,
even though the Start/
Stop button (A) is
pressed.
This error code
is displayed when
the measured blood
pressure value is
excessively low or high.
This error code is
displayed when there
is a fault in the air
circulation of the device.
This error code
is displayed when
too much pressure is
applied to the cuff.
This error code
is displayed when
there is a fault during
measurement.
When the error code "EP" is displayed, please contact the manufacturer.
If the fault persists, please contact the manufacturer before trying to repair the device
on your own.
Cause
The batteries are empty.
The batteries are not inserted
correctly / the polarity (+/-)
is wrong.
The cuff is not attached
correctly.
You talked during
measurement.
You moved yourself or your
arm during measurement.
The cuff is not connected
correctly with the device.
The pressure in the cuff
exceeds 300mmHg.
No measured valued can be
determined.
31
Corrective action
Replace the batteries by new
ones.
Insert the batteries correctly.
Attach the cuff correctly.
Repeat the measurement and
avoid moving and talking.
Check all connections between
the cuff and the device and
repeat the measurement.
Switch off the device and
repeat the measurement.
Switch off the device, attach
the cuff correctly and repeat
the measurement.