HP-IB Error: Not Enough Memory Available for Save.
This message will be generated when the control program tries to save the current Test Set
state into a Save/Recall register using the REG:SAVE commands, but there is insufficient
memory available in the Test Set. The Test Set's non-volatile RAM is shared by the
following resources:
IBASIC programs
Save/Recall registers
RAM Disk
In order to save the current Test Set state into a Save/Recall register more non-volatile
RAM will have to be made available. This can be done by,
reducing the size of the IBASIC program
deleting one or more existing Save/Recall registers
recovering RAM Disk space
The ROM Disk utility RAM_USAGE will display the total amount of non-volatile RAM
installed in the Test Set, the RAM Disk allocation, the Save/Recall register allocation and
the amount of non-volatile RAM available to IBASIC.
HP-IB Error: Unknown Save/Recall error.
This error can occur on a Test Set when trying to SAVE the instrument state to a mass
storage device with a LIF formatted media. The default file system for the Test Set is
DOS. Refer to <blue>"Default File System" on page 346 for more details.
HP-IB Error: HP-IB Units cause invalid conversion of attr.
This error is generated when trying to change Attribute Units and one of the Data Function
values is set to zero. If this error is encountered the programmer must change the Data
Function settings to values that can be converted to the new units_of_measure. Refer to
<blue>"Changing Attribute Units." on page 75 for more details.