The LOCAL key returns the Test Set to local, front-panel control. The Test Set
returns to Local operation (full front-panel control) when either the Go To Local
(GTL) bus command is received, the front-panel LOCAL key is pressed or the
REN line goes false. When the Test Set returns to local mode the output signals
and internal settings remain unchanged, except that triggering is reset to
TRIG:MODE:SETT FULL;RETR REP. The LOCAL key will not function if the
Test Set is in the local lockout mode.
MEAS RESET
The MEAS RESET key clears the measurement history for all of the Test Set's
measurement algorithms: Averaging (AVG key, Spectrum Analyzer trace
averaging), Measurement limit checking (HI LIMIT and LO LIMIT keys), Peak
Hold (AF Analyzer peak hold detectors, Spectrum Analyzer trace peak hold),
autotuning and autoranging, and re-starts all active measurements. The HP-IB
commands :MEASure:RESet are used to select this function programmatically.
Syntax
:MEASure:RESet
Example
OUTPUT 714;":MEAS:RES"
This resets all of the active measurements in the Test Set.
PRESET
The PRESET key resets the Test Set to its power-up state. The IEEE 488.2
Common Command *RST is used to select this function programmatically.
Syntax
*RST
Example
OUTPUT 714;"*RST"
This resets the Test Set to its power-up state.
Equivalent Front-Panel Key Commands