Chapter 1, Using HP-IB
IEEE 488.1 Remote Interface Message Capabilities
Table 4
Test Set IEEE 488.1 Interface Message Capability (Continued)
Message Type
Trigger
Take Control
Abort
44
Implemented
Yes
If in remote programming mode and addressed to listen,
the Test Set makes a triggered measurement following the
trigger conditions currently in effect in the instrument.
The Test Set responds equally to the Group Execute Trig-
ger (GET) bus command or the *TRG Common Com-
mand.
Yes
The Test Set begins to act as the Active Controller on the
bus.
Yes
The Test Set stops talking and listening
Response
IEEE
488.1
Message
GET
MLA
TCT
MTA
IFC