Chapter 8, Programming The Call Processing Subsystem
Programming The CALL CONTROL Screen
SCM:
SID
506
This field displays the decoded station class mark information received from the
mobile station on the reverse control channel in response to a forward control
channel message. The decoded SCM consists of: the mobile station power class
(Class I, II, or III), the transmission type (continuous/discontinuous), and the
transmission bandwidth (20 MHz or 25 MHz).
The
field is only displayed when the
SCM:
reverse control channel message has been decoded.
Refer to the
Display field description, on page 483
read data in the upper right hand portion of the
This field is used to set the system identification number (SID) of the Test Set.
The :SIDentify command is used to control this field.
The
field is an immediate action field. That is, whenever the :SIDentify
SID
command is sent, the change is reflected immediately in the appropriate signaling
message(s) being sent on the forward control channel. No change occurs to the
current state (i.e. Active, Register, Page, Access, Connect) of the Call Processing
Subsystem.
The query form of the command (that is, :SIDentify?) can be used to determine
the current system identification number (SID) setting.
Syntax
:SIDentify <integer number>
:SIDentify?
Example
OUTPUT 714;"CALLP:AMPS:SID 231"
OUTPUT 714;"CALLP:AMPS:SID?"
ENTER 714;Sid_number
field is set to
Display
, for information on how to
screen.