Data Log Setting - Thermo Scientific Orion Star A320 Series Manual

Electrochemistry portable meters
Table of Contents

Advertisement

Thermo Scientific
Timed
Measurement values are continuously updated on the display and the Stabilizing or Ready
icon indicates the measurement stability status. When the data log and export data settings are
enabled, measurements are saved to the data log and exported to a printer or computer at the
selected time intervals from 3 seconds (00:00:03) to 24 hours (24:00:00).
To Set Measurement Read Type:
1. In the measurement mode, press the setup key to access the main setup menu.
2. Press the p, q, t or u key to highlight pH Channel, pH/ISE Channel, COND Channel or
DO/RDO Channel and press the f3 (select) key.
3. Press the p or q key to highlight Mode and Settings and press the f3 (select) key.
4. Press the p or q key to highlight Read Type and press the f3 (select) key.
5. Press the p or q key to highlight Auto, Continuous or Timed and press the f3 (select) key
to set the desired measurement mode.
a. If Timed is selected, press the t or u key to highlight the hours (00:00:00), minutes
(00:00:00) or seconds (00:00:00).
b. Press the f3 (edit) key to access the popup number entry screen.
c. Press the p, q, t or u key to highlight a number, press the f3 (enter) key to select
the number and repeat until a two digit value is shown at the top of the screen.
d. Press the f2 (done) key to save the value and exit the number entry screen.
e. Repeat until the correct time interval is entered for the hours, minutes and seconds.
Time intervals from 3 seconds (00:00:03) to 24 hours (24:00:00) can be entered.
6. Press the f1 (back) key to navigate out of the menu and then press the measure (esc) key
to return to the measurement mode.

Data Log Setting

To save up to 5000 measurement sets to the meter data log, turn the data log setting on in the
setup menu.
1. In the measurement mode, press the setup key to access the main setup menu.
2. Press the p, q, t or u key to highlight Settings and press the f3 (select) key.
3. Press the p or q key to highlight Data Log and press the f3 (select) key.
4. Press the p or q key to highlight On and press the f3 (select) key.
5. Press the f1 (back) key to navigate out of the menu and press the measure (esc) key to
return to the measurement mode.
Chapter 7
Data Transfer and Software Updates
|
Orion Star A320 Series Portable Meter User Manual
77
|

Advertisement

Table of Contents
loading

Table of Contents