Arjohuntleigh Enterprise E5000X Instructions For Use Manual page 58

Table of Contents

Advertisement

Guidance and manufacturer's declaration – electromagnetic immunity
The bed is intended for use in the electromagnetic environment specified below. The
customer or the user of the bed should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
± 6kV contact
discharge (ESD)
± 8kV air
IEC 610000-4-2
± 2kV for power
Electrical fast
supply lines
transient/burst
IEC 61000-4-4
± 1kV for
input/output lines
± 1kV line to line
Surge
IEC 61000-4-5
± 2kV line to
earth
70% U
(30% dip in U
for 25 cycles
40% U
Voltage dips, short
(60% dip in U
interruptions and
for 5 cycles
voltage variations
on power supply
<5% U
input lines
(>95% dip in U
IEC 61000-4-11
for 0.5 cycle
<5% U
(>95% dip in U
for 5 sec
Power frequency
(50/60Hz) magnetic
3A/m
field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to the application of the test level
T
746-577-UK-6
Compliance
level
± 6kV contact
± 8kV air
± 2kV
Not applicable
± 1kV
± 2kV
T
)
T
70% of input
voltage for 500ms
T
)
40% of input
T
voltage for 100ms
<5% of input
T
)
voltage for 10ms
T
<5% of input
voltage for 5000ms
T
)
T
3A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the bed requires continued
operation during power mains interruptions,
it is recommended that the bed be powered
from an uninterruptible power supply or
battery.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
57