Function lockout

When a function is locked, any associated functions are
automatically disabled, e.g. locking the backrest also disables
Bio-Contour and Auto-Chair.
Function lockout settings are retained if the bed is
disconnected from the electricity supply.
746-577-UK-6
Function lockout can be used to prevent operation of the
controls, e.g. when inadvertent movement of the mattress
platform could injure the patient.
The indicator above each function shows its status: on =
function locked; off = function unlocked.
To lock (prevent) ALL functions: press and hold the
Function Lock button for five seconds. The "lock" indicator
above each function button will light.
To selectively lock (prevent) functions:
Press the Function Lock button. The indicator above the
button will light.
Press the ACP button(s) corresponding to the function(s)
to be locked. The "lock" indicator above the button will
light.
After five seconds, the indicator above the Function Lock
button will go out and the lockout settings are stored.
To unlock (allow) previously locked functions:
Press the Function Lock button. The indicator above the
button will light.
Press the ACP button(s) corresponding to the function(s)
to be unlocked. The "lock" indicator above the button will
go out.
After five seconds, the indicator above the Function Lock
button will go out and the lockout settings are stored.