Recharging the
backup battery
WARNING
The battery must only be recharged using the built-in charger. Do
not use a separate charger or power supply.
The backup battery must be ventilated while recharging. Do not
cover the battery or obstruct the area around it.

Duty cycle lockout

746-577-UK-6
To recharge the battery, connect the bed to the electricity
supply. Allow at least eight hours to recharge the battery
when it is completely discharged.
While the battery is recharging, the ACP battery indicator
lights yellow. The indicator will go out when the battery is
fully charged.
Location of backup battery
Continuous operation of the controls may exceed the duty
cycle of the bed's electrical system, causing the indicators
above the buttons to flash. After 30 seconds, the
indicators will light and all functions are locked.
If this happens, wait for at least 18 minutes then follow the
unlocking procedure described in the section Function
lockout.