6
Troubleshooting and Diagnostics
Pump Leak Rate Test

Troubleshooting the Pump Leak Rate Test

Secondary Leak > 3 µL/min
Dynamic Leak > 30 µL/min or Dynamic Leak Rate Test fails
1290 Infinity II High-Speed Pumps User Manual
Leak between the OBV and automatic purge valve
Check for visible leaks on fittings and connectors
Connector not fixed / tight enough
Connector damaged
Leaky filter frit assembly
Remove the seal wash tubes from the support ring and check for leak into
the seal wash path
Main seal leaking/damaged
Piston damaged
Remove waste lines from the automatic purge valve
Damage to rotor seal and/or stator head
Outlet valve not properly assembled
Re-tighten the outlet valve
Check the position of the gold seal
Air in the primary pump chamber
Check for air in the solvent inlet lines and the Tuning signal
Purge the lines, condition and if necessary prime the pump head
Abort due to over pressure
Check solvent and solvent settings
Purge and condition the pump head with water
Leak in inlet valve
Check for moving air bubbles in tubing directly to the inlet valve
Purge the lines with water to remove dirt
Knock at the valve, clean it or replace it
102