System Pressure Test

System Pressure Test

The System Pressure Test is performed to evaluate the leak tightness of the
system up to the point where the system is capped off. The test is started from
the Services & Diagnostics section in the Agilent Lab Advisor Software or in the
Local Controller. Modules such as pump, sampler, column compartments as well
as accessories like valves or columns can be included into the flow path for this
test.
Preparations:
1 Flush the system with HPLC grade water for several minutes from any solvent
2 Start the System Pressure Test and choose the pressure you want to test the
3 Choose the channel with HPLC grade water and if you want to include or skip
4 Click OK and follow the instructions: Place a blank nut into the port up to
Evaluation:
The result as well as the applied limits are displayed after the automatic
evaluation. The limits are:
A report can be displayed, saved or printed by opening it with the Print Result
button at the lower right of the screen.
If the test does not pass, check the system for leaks or call a local Agilent
representative.
Figure 11
1290 Infinity II High-Speed Pumps User Manual
channel.
system with. Consider pressure limits of modules or accessories included
into the flow path.
an additional purging step.
which you want to test the leak tightness of the system.
The test runs automatically without any further user interaction.
For a pressure setting ≤1000 bar, the allowed leak limit is ≤5 µL/min
For a pressure setting >1000 bar, the allowed leak limit is ≤15 µL/min
on page 105 shows a typical test run.