Pump Leak Rate Test
Evaluation:
The result as well as the applied limits are displayed after the automatic
evaluation. The limits are:
•
•
A report can be displayed, saved or printed by opening it with the Print Result
button at the lower right of the screen.
If the test does not pass, check the system for leaks or call a local Agilent
representative.
Figure 9
Figure 9
1290 Infinity II High-Speed Pumps User Manual
The allowed leak limit for the secondary piston is ≤3 µL/min
The allowed leak limit for the primary piston is ≤30 µL/min
on page 100 and
Static (secondary) Leak Test
Figure 10
on page 101 show a typical test run.