Differential Measurement - Ahlborn ALMEMO 2490-1A Operating Instructions Manual

Universal measuring instruments
Table of Contents

Advertisement

The 1st time key
The 2nd time key
To cancel adjustment press key
3. Gain adjustment
For gain adjustment the gain calibration resources must be provided
(see Table). Gain adjustment is then performed by pressing key
exactly the same way as for zero-point adjustment.
 
If correction val. have been programmed, the CORR arrow lights up.
4. Deleting adjustment values
Adjustment values can be cleared by pressing and holding down the key
. On pH probes you can thus restore the default values, base value 7.00
CLR
and gain -0.1689.
Temperature compensation
On conductivity probes and O
temperature compensation is performed automatically. For pH probes a temper-
ature sensor can be specially configured for this purpose; (see Man. 6.2.6).

Differential measurement

10.6
On version 2490-2, if two sensors with the same units and same decimal point
position are connected at measuring points M0 and M1, the difference M1 - M0
appears automatically below the measuring point M2. The sensors are electri-
cally isolated by means of photovoltaic relays. If the differential channel is not
required, it must be cleared via the interface. If further differential channels are
needed, these can also be created via the interface using the appropriate refer-
ence channels; (see Manual, Section 6.3.4).
is pressed the setpoint flashes in the display.
CLR
is pressed, adjustment is performed.
CLR
M▲
probes with an integrated temperature sensor
2
10. Measuring operations