5. Introduction
characteristic, e.g. 10 thermocouple types, NTC and PT100 probes, infrared
sensors, and flow transducers (rotating vanes, thermoanemometers, Pitot
tubes). For humidity sensors additional function channels are available for cal-
culating humidity variables such as dew point, mixture ratio, vapor pressure, and
enthalpy. Even complex chemical sensors are supported. Measured values from
other sensors can also be acquired using the voltage, current, and resistance
ranges with individual scaling in the connector. Existing sensors can also be
used - so long as the appropriate ALMEMO ® connector is connected via its
screw terminals. For digital input signals, frequencies, and pulses, adapter con-
nectors are available with an integrated microcontroller. It is thus possible to
connect virtually any sensor to any ALMEMO ® measuring instrument and to
change sensors without the need for any extra settings.
Function channels
Maximum, minimum, average, and differential values from certain measuring
points can be programmed as function channels, also internal channels, and
can be processed like normal measuring points.
Units
The 2-character units display can be adapted for each measuring channel so
that both the display and via interface always indicate the correct units, e.g.
when a transmitter is connected. Conversion between °C (Centigrade) and °F
(Fahrenheit) is performed automatically.
Measured value designation
Each sensor is identified by means of a 10-character alphanumeric name. This
name is entered via the interface and will appear via interface and in the soft-
ware.
Correction of measured values
The measured value on each measuring channel can be corrected both in terms
of zero-point and gain; this means that even sensors usually requiring initial ad-
justment (e.g. expansion, force, pH) can be freely interchanged. Zero-point cor-
rection and, partly at least, gain adjustment can be performed at the touch of a
button. Sensors with multi-point calibration can also be connected; (see Manual
Section 6.3.13).
Scaling
The corrected measured value on each measuring channel can also be further
scaled in terms of zero-point and gain - using the base value and factor. The dec-
imal point position can be set by means of the exponent function.
Limit values and alarm
Per measuring channel two limit values can be set (1 maximum and 1 minimum).
In the event of one of these limit values being exceeded relay output modules
actuate the associated alarm contacts; these can be allocated individually to
specific limit values. Hysteresis is set by default to 10 digits but this can be ad-
justed to any number between 0 and 99. The exceeding of a limit value can also
be used to start or stop measured value recording automatically.
10