SECURITY SEAL INSTALLATION
If your metrology laws require the use of physical sealing, a lead and wire security seal can be
installed on the MV2 to prevent the end cap from being removed from it to gain access to the
calibration switch, thereby preventing unauthorized access to the calibration adjustments.
To install a lead and wire security seal, refer to
Figure No. 13 and follow the steps provided below:
1. Make certain the two drilled screws securing
the left end cap (as viewed from the MV2 front)
are securely tightened before proceeding.
2. Thread the sealing wire through the hole in the
drilled screws.
3. Pull the wire tight and install the lead seal.
4. The screws cannot be removed without
damaging the seal.