U U Sb PORT
With USb PORT displayed, press the ENTER key. The display will change to USb MOdE.
Proceed to the USb MOdE (USB Port Output Mode) parameter.
Otherwise, press the ID/HEIGHT key to advance to the
NOTE: If the Redbird Wi-Fi and Bluetooth Low Energy (BLE) wireless transmitter module is
installed, pressing the ID/HEIGHT key will advance to the Bluetooth/Wi-Fi parameters with
REdbIRD REV
the display showing
the top of the display. Refer to the Setup and Operation Manual for MedVue Medical Weight
Analyzer with Wi-Fi and Bluetooth, 8555-0550-0M.
USb MOdE (USB Port Output Mode)
The display will show USb MOdE and the current setting. If the value displayed is acceptable,
press the ENTER key to save it. Otherwise use the numeric keys to enter a new value, and
then press the ENTER key to save it.
Allowable values are 0, 1, 2, 3, or 4.
0 = USB Port Output Mode is Disabled
1 = USB Port Output Mode is set for Weight on Demand
(requires serial driver .inf file for Windows)
2 = USB Port Output Mode is set for Continuous Data
(requires serial driver .inf file for Windows)
3 = USB Port Output Mode is set for a Printer
(only works for printers with an internal font)
4 = USB Port Output Mode is set for Welch Allyn Operation
NOTE: Requires device to have an activated license. Must purchase
Detecto WACONNECT.
PRINTER TYPE (Set Type of Printer Output –
The display will show PRINTER TYPE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0, 1, 2 or 3.
0 = Print default ticket
1 = Print journal line
2 = Print label
3 = Print nControl ticket (must have loaded a ticket format, otherwise will use
default ticket)
ENd OF PRINT (End of Print Linefeeds –
At the end of data sent to a printer, the MV2 can send a number of carriage return/line feeds
to space the paper in the printer to a desired position for withdrawal or for the next print.
The display will show End OF PRINT and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0 through 9.
CALIBRATION SETUP, CONT.
and the Redbird Software version in large numbers at
8555-0530-0M Rev E MedVue MV2 Owner's
TIME SETTINGS
Only if USb MOdE = 3
)
Only if USb MOdE = 3
menu.
)