Body Mass Index (Bmi) Operation With Dhr; Using Digital Height Rod - Detecto MedVue MV2 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

Body Mass Index (BMI) Operation with DHR

Using Digital Height Rod

IMPORTANT! Before performing the operations on this page, the digital
height rod must be in its "starting position" (inner sliding tube must be
down completely inside outer stationary tube and headpiece folded flat
against stationary tube) prior to turning on the MV2.
1. Make sure height rod is in starting position.
2. Press the ON/OFF key to turn MV2 on.
3. Press the ZERO key to zero weight display. Note that height reading will show
110.0 cm or 1100 mm (depending on the HEIGHT UNITS setting in Setup).
4. Perform tare operation if required. Refer to Push Button Tare or Tare Weight Entry in
previous section for instructions on using tare.
NOTE: If patient is shorter than 110 cm (1100 mm) or taller than 201.2 cm
(2012 mm), proceed to step 5 or 6 depending on the HEIGHT UNITS setting.
Otherwise, press the ENTER key without entering a value for steps 5 or 6 and
proceed to Step 7.
5. Press ID/HEIGHT key. If Centimeters was selected for HEIGHT UNITS in setup, the
display will change to show CENTIMETERS.
a. Use numeric keys to enter up to 4 digits (# # # . #) for height in centimeters and
then press the ENTER key.
6. Press ID/HEIGHT key. If Millimeters was selected for HEIGHT UNITS in setup, display
will change to show MILLIMETERS.
a. Use numeric keys to enter up to 4 digits (# # # #) for height in millimeters and then
press the ENTER key.
7. Assist patient onto scale.
8. Grasp height rod at hinge pin (see Figure No. 9) and
raise it well above patient's head, and then lift
headpiece to horizontal position. Carefully lower
height rod until headpiece rests on patient's head.
9. BMI will automatically be displayed with height.
10. When weight is stable (STABLE
turn on), read weight, height, and BMI displays.
11. If a printer is connected to scale, press the PRINT
key to print a ticket.
12. While holding headpiece horizontal, raise height rod
well above patient's head.
13. Assist patient off scale and return height rod to
starting position.
14. If Tare operation was performed, press the
NET/GROSS key to return to Gross weight mode.
15. Press the ZERO key. This will reset tare weight to zero.
16. BMI display will clear when weight returns to zero.