Calibration Setup (Non-Protected Setup Parameters) - Detecto MedVue MV2 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

S S YSTEM
With SYSTEM displayed, press the ENTER key. The display will change to LANGUAGE. Proceed
to the LANGUAGE (Language) parameter.
Otherwise, press the ID/HEIGHT key to advance to the
LANGUAGE (Language)
The display will show LANGUAGE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0, 1, 2, or 3.
0 = English
NOTE: The default language is English. Changing the language at this
prompt will change all subsequent prompts and displays to the language
selected.
bACkLIGHT (Backlight Intensity)
The display will show bACkLIGHT and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0 through 255.
0 = Backlight OFF
NOTE: The backlight intensity setting will affect battery life. A lower intensity
setting will conserve power and maximize battery operation.
kEYPAd bEEP (Allow Keypad Beep)
This setting determines whether the MV2 beeps each time a key is pressed.
The display will show kEYPAd bEEP and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise, use the numeric keys 0/NO or 1/YES
to select a new setting, and then press the ENTER key to save it.
Allowable values are YES or no.
YES = The MV2 beeps each time a
key is pressed.
32
CALIBRATION SETUP, CONT.
(Non- Protected Setup Parameters)
1= Spanish
through
8555-0530-0M Rev E MedVue MV2 Owner's
SERIAL PORT
2 = French
3 = German
255 = Backlight at full intensity
no = MV2 does not beep when a key
is pressed.
menu.