H H I RESOLUTION (High-Resolution Mode)
The High-Resolution Mode feature will display the weight at 10 times the standard
resolution. For example, a MV2 configured for 270 kg x 0.1 kg will display weight in high
resolution mode at 270 kg x 0.01 kg. NOTE: If dECIMAL=3, the High-Resolution Mode is
not available.
The display will show HI RESOLUTION and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise, use the numeric keys 0/NO or
1/YES to select a new setting.
Allowable values are YES or no.
YES = If High-Resolution Weight
mode is desired
If viewing the High-Resolution Weight mode immediately is desired:
1. With HI RESOLUTION displayed, press 1/YES key.
2. Press the ENTER key.
3. The display will change to RS232 MOdE.
4. Press the PRINT key twice.
5. The MV2 will reset and display the weight in High Resolution mode.
Otherwise, after selecting the value for the High-Resolution mode parameter, press the
ENTER key to save the selection and advance to the SERIAL PORT, RS232 MOdE (RS232
Port Output Mode) parameter to continue Calibration Setup.
CALIBRATION SETUP, CONT.
8555-0530-0M Rev E MedVue MV2 Owner's
no = High-Resolution Weight
mode is not desired
35