Detecto MedVue MV2 Owner's Manual page 42

Medical weight analyzer
Table of Contents

Advertisement

S S ERIAL PORT
With SERIAL PORT displayed, press the ENTER key. The display will change to RS232 MOdE.
Proceed to the RS232 MOdE (RS232 Port Output Mode) parameter.
Otherwise, press the ID/HEIGHT key to advance to the
RS232 MOdE (RS232 Port Output Mode)
The display will show RS232 MOdE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0, 1, 2, or 3.
0 = RS232 Port Output Mode is Disabled
1 = RS232 Port Output Mode is set for Weight on Demand
2 = RS232 Port Output Mode is set for Continuous Data
3 = RS232 Port Output Mode is set for a Printer
bAud RATE (Serial Output Baud Rate –
The display will show bAud RATE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 1 through 9, and 0.
1 = 2400 Baud
2 = 4800 Baud
3 = 9600 Baud
PRINTER TYPE (Set Type of Printer Output –
The display will show PRINTER TYPE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0, 1, 2, or 3.
0 = Print default ticket (P185 or other type of ASCII printer)
1 = Print journal line
2 = Print label (P220)
3 = Print nControl ticket (must have loaded a ticket format, otherwise will use
default ticket)
ENd OF PRINT (End of Print Linefeeds –
At the end of data sent to a printer, the MV2 can send a number of carriage return/line feeds
to space the paper in the printer to a desired position for withdrawal or for the next print.
The display will show End OF PRINT and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a new
value, and then press the ENTER key to save it.
Allowable values are 0 through 9.
36
CALIBRATION SETUP, CONT.
4 = 14.4k Baud
5 = 19.2k Baud
6 = 28.8k Baud
8555-0530-0M Rev E MedVue MV2 Owner's
USb PORT
menu.
Only if RS232 MOdE = 1, 2 or 3
7 = 38.4k Baud
8 = 56k Baud
9 = 57.6k Baud
0 = 115.2k Baud
Only if RS232 MOdE = 3
Only if RS232 MOdE = 3
)
)
)