Interconnections; Optional Ac Power Adapter; Height Rod Option - Detecto Dhr; Load Cell - Detecto MedVue MV2 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

INTERCONNECTIONS

All input, output, power and option connections to the MV2 are made on the its rear panel.
Connections for the Load Cell input and the Serial I/0 are made using 9-pin "D" shaped
connectors. The Height Rod option (Detecto DHR, Digital Height Rod) and Ethernet (MVIP)
options use snap-in modular connectors. The USB port is a device (or upstream) port using an
industry standard "B" connector. The optional 12VDC wall plug-in UL/CSA listed AC power
adapter is connected using a power jack. Refer to Figure No. 1.

Optional AC Power Adapter

To power the MV2 using the optional 12VDC wall plug-in AC power adapter, connect the
plug from the adapter into the power jack on the back of the MV2, and then plug the power
adapter into the proper electrical outlet. Refer to Figure No. 1. On models requiring 220
VAC, it is the customer's responsibility to obtain the correct power adapter plug.
Height Rod – Detecto DHR (Digital Height Rod) Option
The optional Detecto DHR (Digital Height Rod) connects to the MV2 by a snap-in modular
connector socket. Insert the modular connector of the DHR cable into the socket on the rear
of the MV2 it locks in place (a clicking sound will be heard when it is locked in place).

Load Cell

The load cell cable connects to the MV2 via a 9-pin "D" connector on the rear panel of the
MV2. Figure No. 4 shows the pin identification for the load cell connector. Make certain that
the pins are correctly identified before soldering a wire to them. Use the connector retaining
screws to hold the load cell cable connector securely to the rear panel.
PIN NO.
FUNCTION
1
+ EXCITATION
2
- SIGNAL
5
SHIELD
6
- EXCITATION
7
+ SIGNAL
NOTE: Pins 3, 4, 8, and 9 are not used.
DESCRIPTION
CONNECTOR
CONNECTOR SHELL

USB

The USB port on the MV2 is a device (or upstream) port and uses readily available cables
with the industry standard "B" connector.
USB, Serial, Ethernet, and Bluetooth/Wi-Fi Interface Specifications
The USB, Serial, optional Ethernet interface, and Bluetooth/Wi-Fi on the Model MV2C can
be configured during the setup and calibration procedure or during the setup review
operation. Using either method, it is possible to select the operation of the interfaces.
The interfaces may be connected to a computer for transmission of weight and associated
data to a PC-based EMR (electronic medical record) software program. The data can be
transmitted on demand (pressing the PRINT key) or on receipt of a command from the
computer.
INSTALLATION, CONT.
MATING CONNECTOR INFORMATION
ITEM
DE9-P
C883010001
8555-0530-0M Rev E MedVue MV2 Owner's
5
4
3
2
9
8
7
Figure No. 4
CARDINAL PART #
6610-2379
6610-1131
1
6
5

Advertisement

Table of Contents
loading

Table of Contents