11.
Align the MINILOADER and PROCESSOR horizontally as shown in figure 16, fit the other two M4
LARGE DIAMETER WASHERS and the M4 NUTS. Do not tighten the NUTS yet. It is very important
to have the alignment correct, especially for an RA PROCESSOR because the CHUTE EXTENSION can
interfere with the LED's inside the FILM ENTRY.
Figure 16
HORIZONTAL
ALIGNMENT
12.
By adjusting the LEVELLING SCREWS on the MINILOADER, (and the PROCESSOR if necessary),
set the height of the MINILOADER. The vertical alignment is correct when a FILM hits the lower EN-
TRY ROLLER 1.5 mm below the nip of the ROLLERS. See figure 17.
Figure 17
VERTICAL
ALIGNMENT
13.
When both horizontal and vertical alignments are correct, tighten the M4 NUTS.
14.
Go to the INSTALLATION CHECK LIST on page II-22 of this publication to complete the installation..
December 1992
PLAN VIEW
MINILOADER
Film
1.5 mm
Page II-15
PROCESSOR
Align the two frames here.
Entry rollers
of processor
© Kodak Ltd.
II 3239