INSTALLATION INSTRUCTIONS
14.
Check that the PROCESSOR (if fitted) is brought out of stand-by when a CASSETTE is entered in the
MINILOADER.
15.
Set the date and time on the MICROPROCESSOR (see the THEORY GUIDE). Zero all statistics and
film counters (operator and ESR).
16.
Remove the STEP-BY-STEP SWITCH and the SERVICE KEY and re-install all the PANELS.
17.
Load the MINILOADER 2 PLUS with some customer FILM and make some test cycles. Examine the
processed film for artifacts such as scratches, SUCKER marks and light fog. Correct as necessary. If the
MINILOADER is a STAND-ALONE unit, if possible take the film elsewhere to process and do the
checks.
18.
Fill in the INSTALLATION REPLY FORM, and note the meter (counter) reading. Make sure the correct
SERVICE CODE is used when feeding back the installation to SCAN.
SERVICE CODE
3236
3239
3240
December 1992
MODEL
STAND-ALONE
M35-M
480RA
Page II-25
II 3239
© Kodak Ltd.