7.
Functional Units
7.18 Distillation
The distillation systems consists of the still with ascending pipe, condenser, water separator, flash
tank, extraction tank and vacuum pump. The still is equipped with an overfill preventer. The still and
the sight-glass are rinsed each time the still is filled. You can turn the distillation heater on and off
with 13E. A thermal sensor at the bottom turns off the heating when a selected value has been
exceeded. Distillation takes place at an underpressure of approximately 900 mbar (13 psi).
Besides you can stop the destillation by pressing the button „0" (about 2 sec.)
1 Still with ascending pipe
2 Condenser (stainless steel)
3 Overfill sensor
4 Flash tank
46
5 Extraction tank
6 Vacuum pump
7 Water separator
7.
7.18
707766-12-0
M12/M15/M18