Setting Up The Cable Entry (Only For A Device With A Manual Analytical Draft Shield); Acclimatization - Sartorius Cubis MCE series Operating Instructions Manual

Semi-micro-, analytical and precision balances
Hide thumbs Also See for Cubis MCE series:
Table of Contents

Advertisement

5.9
For models with a manual analytical draft shield, a cable can be fed into the
weighing chamber, e .g . when using a temperature sensor .
Procedure
t
t
2
1
1
t
t
t
1
t
t
1

5.10 Acclimatization

When a cold device is brought into a warmer area: The temperature difference can
lead to condensation of humidity in the device (moisture formation) . Moisture in the
device can lead to malfunctions .
t
Setting Up the Cable Entry (Only for a Device with a
Manual Analytical Draft Shield)
Lift the locking tab (1) on the rear panel of the device .
Lift the panel (2) out of the device .
Rotate the panel 180° so that the recess (1) in the panel points towards the
weighing module .
Feed the connection cable into the weighing chamber .
Insert the panel into the guide groove (1) .
Lift the locking tab (1) on the rear panel of the device and push down the panel .
Press the locking tab down and close it .
Allow the device to acclimatize for approx . 2 hours at the installation site .
Ensure that the device is disconnected from the power supply during that time .
Installation
Cubis
®
MCE Operating Instructions
33