7.2 Wiring

10
TSX -20°C Flammable Materials Storage Laboratory Freezers
The wiring diagrams are attached on the back of the cabinet.
CAUTION: Connect the equipment to the correct power source. Incorrect
voltage can result in severe damage to the equipment.
CAUTION: Risk of Shock. For personal safety and trouble-free operation,
this unit must be properly grounded while in use. Failure to ground the
equipment may cause personal injury or damage to the equipment. Always
conform to the National Electrical Code and local codes. Do not connect
the unit to overloaded power lines.
Do not position the unit in a way that impedes access to the disconnecting
device or circuit breaker in the back of the unit.
Always connect the unit to a dedicated (separate) circuit. Each unit is
equipped with a service cord and plug designed to connect it to a power
outlet which delivers the correct voltage. Supply voltage must be within
±
10% of the unit rated voltage. If cord becomes damaged, replace with a
properly rated power supply cord.
Table 3. Power Cord Specifications
Model
Power Cord Specifications
A
3-G 12 AWG, NEMA 5-15P, 15 A/125 V
CAUTION: Risk of Shock. Never cut the grounding prong from the
service cord plug. If the prong is removed, the warranty is invalidated.
In an emergency, the power cord is a disconnect device.
The -20°C Flammable Materials Storage Laboratory Freezers are equipped with a
connectivity port on the back of the unit. For use, refer to the instructions
provided with the applicable Thermo Fisher Scientific connectivity device.