Installation

5 Installation

5.1 Location

5.2 Wiring

5.3 Electrical
Specifications
6
Installation and Operation Manual
WARNING: Do not exceed the electrical and temperature ratings printed
on the data plate located on the lower left side of the unit.
CAUTION: Improper operation of the equipment could result in
dangerous conditions. To preclude hazard and minimize risk, follow all
instructions and operate within the design limits noted on the dataplate.
Install the unit in a level area free from vibration with a minimum of six
inches of space on the sides, rear, and top. Allow enough clearance so that
the lid can swing open at least 90°.
Do not position the equipment in direct sunlight or near heating diffusers,
radiators, or other sources of heat. The ambient temperature range at the
location must be 59 to 90°F (15 to 32°C).
CAUTION: To allow for proper air flow, a minimum of six inches of
clearance space is required behind the freezer.
CAUTION: Connect the equipment to the correct power source. Incorrect
voltage can result in severe damage to the equipment.
WARNING: For personal safety and trouble-free operation, this unit must
be properly grounded before it is used. Failure to ground the equipment
may cause personal injury or damage to the equipment. Always conform to
the National Electrical Code and local codes. Do not connect the unit to
overloaded power lines.
The models and voltage types are as specified in the following table:
Table 1. Electrical Specifications
Model
Voltage
ULT185-5-A
115V
ULT185-5-V
230V
Frequency
60Hz
50Hz
Thermo Scientific Benchtop Ultra-Low Temperature Freezers
Current
16A, 1Phase
12A, 1Phase