Baskets; Temperature Sensor Bottles (If Present) - Thermo Scientific TSX Series Installation And Operation Manual

-20c flammable materials storage laboratory freezers
Table of Contents

Advertisement

Installation

7.5 Baskets

7.6 Temperature Sensor
Bottles (if present)
16
TSX -20°C Flammable Materials Storage Laboratory Freezers
Baskets can be ordered as an accessory for this unit.
The basket slides are adjustable to higher and lower positions in the cabinet.
You can position these slides in the vertical slots which are spaced at one-inch
intervals.
The baskets must be removed in order to adjust the position of the slides. The
procedure for removing, reinstalling, and changing position of the baskets can be
found in Section 7.4.1, Section 7.4.2, and Section 7.4.3.
The maximum basket capacity is 27 kg (60 lbs).
On some models, the temperature displayed on the control panel is measured by
a probe inserted in sensor bottle inside the cabinet. If present, the sensor probe
must be inserted in the glycol bottle prior to operation.
The sensor bottle is positioned on the back wall of the unit.
To install the sensor(s) into the bottles, complete the following steps:
1. Remove the bottle from the bracket and verify that each bottle is full of
liquid.
2. Remove the solid cap from the bottle and save for the next time the unit
needs to be moved.
3. Remove the taped lid and temperature sensor(s) from the wall.
4. Insert the temperature probe(s) into the bottle.
5. Secure the Lid to the bottle and return the bottle to the bracket.