Mounting Mag 1100 F - Siemens SITRANS F Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Vibrations
Avoid strong vibrations.
Figure 4-7
In applications with strong vibrations, Siemens recommends remote mounting of the
transmitter!
4.6

Mounting MAG 1100 F

The MAG 1100 F sensor has an integrated clamp connection and is designed for installation
between two adapters, which are supplied separately. The adapters are available for a variety
of standards according to: ISO, DIN, SMS, BS and DS, for direct welding into dairy pipes or
with clamp- or threaded fittings.
At assembly, a gasket is located in the recess in the face of the adapter, and the adapter is
then fastened to the MAG 1100 F sensor with a clamp ring.
SITRANS MAG 1100/1100 F
Operating Instructions, 08/2018, A5E03433301-AD
Avoid vibrations
CAUTION
Sensor
Gasket
Process connection adapter
Clamp ring
Installing/Mounting
4.6 Mounting MAG 1100 F
27