Advertisement

Cable specifications
● Only use cables with at least the same degree of protection as the sensor to install the
sensor.
● The line length from the cable gland to the terminals must be kept as short as possible.
Line loops in the terminal box must be avoided.
● To guarantee the IP67 degree of protection, use cables with the required specifications.
5.2

Remote installation

Note
Remote installation only
The following applies to remote installation of MAG 5000 / 6000 or MAG 6000 I.
SITRANS MAG 1100/1100 F
Operating Instructions, 08/2018, A5E03433301-AD
WARNING
Protective conductor terminal
The required cable is min. AGW16 or 1.5 mm
WARNING
Wire insulation
For field wiring installation: Ensure that the national requirements of the country in which
the flowmeters are installed is met.
5.2 Remote installation
Cu.
2
Connecting
35