Removing The Toe/Kick Panels & Service Panels - Labconco SteamScrubber 44003 Series Service – Technical Manual

Glassware washer
Hide thumbs Also See for SteamScrubber 44003 Series:
Table of Contents

Advertisement

Labconco Glassware Washer Service/Technical Manual
Removing the Toe/Kick Panels &
Service Panels (Under-Counter Models)
For many cleaning, maintenance and repair jobs associated with the
FlaskScrubber® and/or SteamScrubber®, the toe & kick panels will need
to be removed.
This will help gain access to parts inside the base of
the washer and is necessary for initial install of the machine.
Remove the Toe/Kick Panel
1. Remove the two screws located at the bottom of the panel.
2. Remove 2 screws on each side of the middle panel.
3. Pull the middle panel forward and remove it.
4. Remove the cover over the electrical junction box by removing the two
screws, disconnecting the wiring connections for the door fan and
sliding the panel out to the left side.
cable and Door Harness.
Removing the Back and Side Service Panels
Locate each of these panels beneath the insulation jacket of the Under-
Counter Models.
There are four bolts holding each panel on.
the bolts and pull off the plates.
When replacing the plates, make sure they are properly aligned and in the
correct orientation on the machine.