Maintaining The Air Gap In The Drain - Labconco SteamScrubber 44003 Series Service – Technical Manual

Glassware washer
Hide thumbs Also See for SteamScrubber 44003 Series:
Table of Contents

Advertisement

Labconco Glassware Washer Service/Technical Manual
* Debris that falls into the sump can damage and/or destroy the wash and
drain pumps.
To Replace the Filter Screen
If the filter screen is damaged or too dirty to clean and needs to be
replaced follow the instructions above to remove the original filter screen
and keep the hardware.
Once removed, position the new screen and insert the
holding screw in place.
Make sure that the 1/8" drilled hole is located
below the water distribution channel.

Maintaining the Air Gap in the Drain

If the installation included an air gap (recommended) it requires
periodic cleaning.
The air gap is not part of the washer; therefore,
it is not covered by the washer's warranty.
The air gap protects the
washer against water backing up in the event of a clogged drain.
Check the air gap for build-up any time your washer is not draining
well.
Most types of air gaps are easy to clean.
Simply follow the manufacturer's
directions for maintaining your air gap.
10
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com