2-10Emergency Recovery

NOTE:
ILLUSTRATION 5-23
RESET RECOVERY SCREEN
om 5184516-100 Rev. 5
If during a procedure power is lost to the system, the system will reboot if power is
returned to the unit after 30 seconds.
Upon reboot, the system will remember all patient parameters if power was lost
during a procedure. If power is off longer than 30 seconds, the system must be
rebooted by pressing the Power On button located on the RCIM. The study will
not resume. Re-select the patient from the Patient list and the exam.
IF
THE
SYSTEM
CONTROL
COMMUNICATION WITH THE REMAINING SYSTEM DURING A
DIGITAL RECORD ACQUISITION, THE USER WILL BE ALLOWED TO
CONTINUE/FINISH
(EXPOSURE
ACQUISITION. ANY SUBSEQUENT ACQUISITION (FLUORO OR
DIGITAL) WILL BE INHIBITED.
The system will then give the options for continuing the study by selecting
Resume Exam or ending the study by the Close Exam option.
INTEGRATED
CONSOLE
SWITCH
IS
RELEASED)
LOSES
THE
5-24