Operation; Parking (Docking) The Imaging Device Carriage - GE Precision 500D Operator's Manual

Table of Contents

Advertisement

SECTION 4

OPERATION

4-1Parking (Docking) the Imaging Device
Carriage
NOTE:
NOTE:
om 5184516-100 Rev. 5
Do not hold onto the power–assist handle during power–up.
If power–up tests fail, repeat the power–up test sequence by pressing the red
RESET button located on the RCIM. If a failure occurs again, call GE service.
1. Verify the emergency STOP button on the table is disengaged.
2. Press the POWER ON button located on RCIM. Power up tests begin.
3. Apply the Lat Long Lock.
4. Disengage the compression lock (Comp Lock).
5. Move the imaging device vertically to maximum height over the table top.
To aid parking, the system disables power assist near maximum vertical
height. Use the grasp handle,
imaging device.
6. Push down the CARRIAGE LOCK LEVER,
simultaneously push the imaging device carriage to the rear. Refer to
TRATION
7-18.
7. At the rear, the CARRIAGE LOCK LEVER will engage in the PARK position.
8. For additional clearance, release the Lat Long Lock and push the imaging
device and table tower assembly further to the rear.
If the carriage is not locked forward, X–ray exposures are inhibited.
(ILLUSTRATION
7-18), to assist positioning the
(ILLUSTRATION
7-18), and
ILLUS-
7-35