4-1Power Line Requirements
TABLE 11-10
MINIMUM WIRE SIZE (65 KW)
PARAMETER
THREE PHASE GENERATOR – 65 KW
Input Voltage
380VAC
Wire Size
Length
15 m (50 ft.)
16 mm2
(#6
30 m (100 ft.)
22 mm2
AWG)
46 m (150 ft.)
30 mm2
(#5
60 m (200 ft.)
44 mm2
AWG)
(#3
AWG)
(#1
AWG)
TABLE 11-11
KVA LOAD CHARACTERISTICS 65KW
Phase
Nominal line Voltage (Vac)
Voltage Range (Vac)
Momentary Line Current (Amp)
Continuous Line Current (Amp)
Power demands (kVA)
Frequency
TABLE 11-12
MINIMUM WIRE SIZE (80 KW)
PARAMETER
THREE PHASE GENERATOR – 80 KW
Input Voltage
380VAC
Wire Size
Length
15 m (50 ft.)
22 mm2
(#5
30 m (100 ft.)
30 mm2
AWG)
46 m (150 ft.)
43 mm2
(#3
60 m (200 ft.)
54 mm2
AWG)
(#1
AWG)
(#0
AWG)
om 5184516-100 Rev. 5
400 VAC
415 VAC
16 mm2
(#6
16 mm2
(#6
22 mm2
AWG)
22 mm2
AWG)
30 mm2
(#5
30 mm2
(#5
44 mm2
AWG)
44 mm2
AWG)
(#3
(#3
AWG)
AWG)
(#1
(#1
AWG)
AWG)
3
3
380
400
+/–10%
+/–10%
147
140
7
6.7
97
97
400 VAC
415 VAC
22 mm2
(#5
22 mm2
(#5
30 mm2
AWG)
30 mm2
AWG)
43 mm2
(#3
43 mm2
(#3
54 mm2
AWG)
54 mm2
AWG)
(#1
(#1
AWG)
AWG)
(#0
(#0
AWG)
AWG)
440 VAC
460 VAC
16 mm2
(#6
16 mm2
22 mm2
AWG)
22 mm2
30 mm2
(#5
30 mm2
44 mm2
AWG)
44 mm2
(#3
AWG)
(#1
AWG)
3
3
420
440
+/–10%
+/–10%
133
127
6.2
6 5.
97
97
47 / 53Hz and 57/63Hz
440 VAC
460 VAC
22 mm2
(#5
22 mm2
30 mm2
AWG)
30 mm2
43 mm2
(#3
43 mm2
54 mm2
AWG)
54 mm2
(#1
AWG)
(#0
AWG)
480 VAC
(#6
16 mm2
(#6
AWG)
22 mm2
AWG)
(#5
30 mm2
(#5
AWG)
44 mm2
AWG)
(#3
(#3
AWG)
AWG)
(#1
(#1
AWG)
AWG)
3
3
460
480
+/–10%
+/–10%
122
117
7
5.5
97
97
480 VAC
(#5
22 mm2
(#5
AWG)
30 mm2
AWG)
(#3
43 mm2
(#3
AWG)
54 mm2
AWG)
(#1
(#1
AWG)
AWG)
(#0
(#0
AWG)
AWG)
11-7