Set the Measurement Unit
11. Select measurement unit with the
Select the moisture content (based on wet
sample ) with the
Set the Minimum Scale Value of the % Display
12. Select the % display with the
Select 0.01 [%] with the
Set the Minimum Scale Value of the Gram Display
13. Select the gram display with the
Select 0.001 [g] with the
Store the Parameters and Finish the Operation
14. Press the
the measurement program to program number 6.
Pressing the key, the weighing mode is automatically
displayed.
When PROG 6 is recalled, the settings can be used.
To cancel the new parameters and return to the
weighing mode, press the
MS-70 / MX-50 / MF-50 / ML-50
or
key.
SELECT
or
key.
or
key to store the new parameters of
ENTER
RESET
key.
SELECT
key.
key.
SELECT
key.
key.
41
Measurement unit
Minimum scale
value of % display
Minimum scale
value of gram
display
160°C
10.0 min
Gram display of
weighing mode
0
" M " O " I " S " T
" %
" / " W
" 0 " 0 " 1
%
" 0 . " 0 " 0 " 1
g
0. 0 00
g