A&D MF-50 Instruction Manual page 40

A&d moisture analyzer instruction manual ms-70 mx-50 mf-50 ml-50
Hide thumbs Also See for MF-50:
Table of Contents

Advertisement

Select an Analysis Mode
5. Press the
SELECT
blink. Select an analysis mode using the
If standard mode is used, select
If automatic mode is used, select
If manual mode is used, select
Set ACCURACY for the Standard Mode
6. Select ACCURACY with the
Select MID. with the
When
ACCURACY
Set the Heating Pattern
7. Select drying temperature 1 with the
key. Select step drying
pattern with the
Set Drying Temperature 1
8. Set 120°C to drying temperature 1 with
the
or
key.
Set the Time
9. Select time with the
Set 5.0 minutes with the
Set Drying Temperature 2
10. Select drying Temperature 2 with
the
SELECT
Set 160°C with the
Set the Preset Termination Value for Automatic Mode
11. Select a preset termination value with the
key. Select 0.05 [%/min] with
SELECT
the
or
key.
Set Measurement Unit
12. Select measurement unit with the
Select moisture content (based on a wet
sample ) with the
key to blink a symbol, it will
U-m
SELECT
or
key.
is selected, HI, MID. or LO blinks.
of the heating
key.
PROGRAM
key.
SELECT
or
key.
key.
or
key.
or
key.
or
key.
.
5td
.
U-a
Automatic mode
.
Manual mode
key.
SELECT
Drying temperature 1
Drying temperature 2
Preset termination value
Standard mode
Manual mode
key.
SELECT
Measurement unit
38
Analysis
" U " - " a
0. 0 00
ACCURACY
ACCURACY
MID.
Step drying
" 1 " 0 " 5 °C
1
" 1 " 2 " 0 °C
Time
" 5 ." 0 min
2
" 1 " 6 " 0 °C
MS-70 / MX-50 / MF-50 / ML-50
g
Standard
mode
Automatic
mode
" 0 ." 0 " 5
%/min
0
" M " O " I " S " T
" %
" / " W

Hide quick links: