A&D MF-50 Instruction Manual page 41

A&d moisture analyzer instruction manual ms-70 mx-50 mf-50 ml-50
Hide thumbs Also See for MF-50:
Table of Contents

Advertisement

Set the Minimum Scale Value of the % Display
13. Select the % display with the
Select 0.01 [%] with the
Set the Minimum Scale Value of the Gram Display
14. Select the gram display with the
Select 0.001 [g] with the
Store the Parameters and Finish the Operation
15. Press the
the measurement program to program number 3.
Pressing the key, the weighing mode is automatically
displayed.
When PROG 3 is recalled, the settings can be used.
To cancel new parameters and return to weighing
mode, press the
Timer Mode
This explanation uses the following parameters and sample displays of MX-50.
Drying temperature 1
"Temperature 1" can be set higher than "temperature 2" in step drying.
Common Items
Program number ..........................................6 ( PROG 6 )
Analysis mode ..............................................Timer mode (Symbol: U-t )
Drying program
Measurement unit ..............................................Moisture content
Minimum scale value during measurement ..0.01 %
Minimum scale value of the gram display.....0.001 g
Sample quantity............................................Approximately 5 g
Data memory function ..................................Not used
MS-70 / MX-50 / MF-50 / ML-50
SELECT
or
key.
or
key to store the new parameters of
ENTER
key.
RESET
Refer to page 37 for Standard Mode, Automatic Mode or Manual Mode
Time 1
Stage 1
Heating pattern....................Step drying (
Drying temperature 1 ..........160°C
Drying temperature 2 ..........120°C
Time 1..................................5.0 minutes
Time 2..................................10.0 minutes
key.
key.
SELECT
key.
Drying temperature 2
Stage 2
39
Minimum scale
value of % display
Minimum scale
value of gram
display
160°C
Gram display of
weighing mode
Time 2
)
% MOIST
D
" 0 . " 0 " 1
%
" 0 . " 0 " 0 " 1
g
0. 0 00
g
/W/D

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ml-50Ms-70Mx-50

Table of Contents