A&D MF-50 Instruction Manual page 62

A&d moisture analyzer instruction manual ms-70 mx-50 mf-50 ml-50
Hide thumbs Also See for MF-50:
Table of Contents

Advertisement

10. After fifteen minutes, the buzzer sounds and the display
blinks
" 1 " 6 " 0 °C
thermometer value using the
11. Press the
the adjustment and to return to the weighing mode.
If the report for GLP, GMP and ISO is output,
is displayed. Output condition is
glp
selected in the function table.
When the heater cover is opened during measurement or
the
STOP
analyzer displays the weighing mode.
Calibration Report Example for Temperature Sensor Adapted to GLP, GMP and ISO
Preset the following parameters to print data to AD-8121B
Device
Interval
Output format
Analyzer
adapted to GLP,
GMP and ISO
AD-8121B
1: Factory settings
...........A & D .........
MODEL......MX-50 .........
S/N.....K1234567 .........
ID.......LAB-123 .........
DATE..2004/09/30 .........
TIME....12:34:56 .........
CALIBRATED
.....TEMPERATURE
TARGET....ACTUAL
100..C.....97..C .........
160..C....162..C .........
SIGNATURE
- - - - - - - -
. Adjust the blinking value to the
key to store the new data, to finish
ENTER
key is pressed, calibration is stopped and the
Parameter
pU5e 1
info 1
or
info 2
MODE 3
}
.........
or
key. (Example: 162°C)
Description
1 Approx. two seconds interval in each line.
Calibration report is printed at "Calibration of the
Weighting Sensor" and "Calibration of Drying
Temperature".
Dump print (Received data is printed as it is)
Manufacture
Model
Serial number
ID number
Date
Time
Calibration type
100°C target value
160°C target value
Signature
60
Buzzer
160℃ " 1 " 6 " 0 ℃
160℃ " 1 " 6 " 2 ℃
end
glp
0. 0 00
0. 0 00
measurement value
measurement value
MS-70 / MX-50 / MF-50 / ML-50
g
g