Pressure/Temperature Testing; Pressure Testing (If Pressure Fitted); Pressure Measurement Good Practice; Large Bore Tubing Issues - UEi C162 Instruction Manual

Flue gas combustion analyzer with heat exchanger test
Hide thumbs Also See for C162:
Table of Contents

Advertisement

Select the Temperature Rotary Dial position.
WARNING
Never attempt to take a pressure reading without knowing the maximum pressure that might be present.
This instrument's pressure transducer is rated at 2.3in H
Using the black connectors and the manometer hose. Connect P1 for a single pressure or P1 and P2 for differential pressure.
C161
WARNING
Before using the analyzer to measure the pressure of a gas/air ratio valve, read the boiler manufacturer's manual thoroughly. If in
If using larger bore tubing when performing pressure tests:
Push tubing over the rim of the
spigot to ensure a gas tight seal.
C161
Find Quality Products Online at:
TEMPERATURE TESTING
Use the T1 connection for the Flue Probe temperature sensor
Use the T2 connection for the Inlet temperature sensor
Real time temperature difference
Real time pressure reading (pressure will only show is fitted)

PRESSURE TESTING (IF PRESSURE FITTED)

0 with a maximum pressure over range of 11.8in H
2
Use the T1 connection for the Flue Probe temperature sensor
Use the T2 connection for the Inlet temperature sensor
Real time temperature difference
Real time pressure reading (pressure will only show is fitted)

PRESSURE MEASUREMENT GOOD PRACTICE

LARGE BORE TUBING ISSUES

GlobalTestSupply
www.
This may not produce a gas tight seal.
.com
sales@GlobalTestSupply.com
0.
2
13