Refer to user guide regarding potential hazard and proper instructions. Before using this meter, read all safety information carefully. In this manual the word "WARNING" is used to indicate conditions or actions that may pose physical hazards to the user. The word "CAUTION" is used to indicate conditions or actions that may damage this instrument.
Higher voltages and currents require greater awareness of physical safety hazards. Before connecting the test leads; turn off power to the circuit under test; set meter to the desired function and range; connect the test leads to the meter first, then connect to the circuit under test. Reapply power. If an erroneous reading is observed, disconnect power immediately and recheck all settings and connections.
This meter is designed for trade professionals who are familiar with the hazards of their trade. Observe all recommended safety procedures that include proper lock-out utilization and the user of personal protective equipment that includes safety glasses, gloves and flame resistant clothing.
| Equipment is safe for connection and disconnection to Live conductors | ![]() | Ground |
![]() | AC Alternating Current | | Warning or Caution |
![]() | DC Direct Current | ![]() | Double Insulation (Protection Class II) |
![]() | Not Applicable to Identified Model | ![]() | Battery |
Measurement Category | Short-Curcuit (typical) kAa | Location in the building installation |
II | < 10 | Circuits connected to mains socket outlets and similar points in the MAINS installation |
III | < 50 | Mains distributions parts of the building |
IV | > 50 | Source of the mains installation in the building |
Always keep your hands and fingers behind the hand guards when measuring current on exposed conductors. Contact may result in serious injury.
Use CAT III listed test leads or higher. DO NOT attempt to measure more than 1000V DC, 750V AC or 2000µA DC.
Displays and Indicators
Location of function icons on display may vary be model number.
![]() | AC indicator |
![]() | DC indicator |
![]() | Indicates a negative value (DC negative voltage) |
MAX | Maximum value displayed |
MIN | Minimum value displayed |
A (top display) | Displaying Amps from clamp jaw or adapter |
ADP (top display) | Displaying value from adapter |
![]() | Low battery indication |
![]() | Hold function |
![]() | Diode function |
![]() | Continuity function |
nF / µF | Capacitance function (nanofarads or microfarads) |
nA | Microamps function (1µA is 0.000001 Amp) |
Hz | Frequency function |
% | Duty cycle function |
mV | Millivolts function (1 mV is 0.001 Volt) |
APO | Auto power off is active |
AT | Auto Range mode is active |
O.L. | Displayed if input value exceeds selected range |
![]() | Resistance function |
°F | Degrees Fahrenheit |
°C | Degrees Celsius |
After powering off, the meter will turn on again when you change ranges, rotate the selector dial or press a button.
NOTE: The APO is disabled during MIN/MAX mode.
Press and hold the HOLD button for 2 seconds to activate. Lights automatically turn off in 2 minutes to save battery life.
NOTE: After activating the lights, quick press to activate the hold mode.
The meter defaults to auto range mode and AT is indicated on the display while active. Press the RANGE button to cycle through available ranges. Selecting a specific range will put the meter in manual range mode and AT will no longer be displayed on screen. Press and hold the RANGE button to return to auto ranging mode.
NOTE: Manual ranging will provide a faster input response over auto range.
When using the MIN/MAX capture mode for amps, it is recommended that you first select the range of the expected maximum value. If this is not done it will lock in the lowest range for the initial measurement. If the maximum value exceeds this range the meter will capture O.L. as the maximum value.
Press the HOLD button to activate. This will freeze the reading and range in the display for review.
NOTE: Max capture is useful for motor inrush current.
NOTE: Frequency greater than 1MHz will display 0.000
DO NOT measure resistance on a live circuit.
Safely discharge capacitor before measurement.
NOTE: Leave meter connected to the capacitor for 10 seconds allowing the reading to settle. Larger capacitors could take up to 60 seconds.
Press SELECT to change between continuity and diode
NOTE: Continuity tone sounds at approximately <50Ω.
NOTE: Shows voltage drop if forward biased and O.L. if reversed biased.
Disconnect test lead probes from voltage source and meter.
NOTE: Leaving clamp head or attachment plugged in will drain battery.
Repair and service of this instrument is to be performed by qualified personnel only. Improper repair or service could result in physical degradation of the meter, altering the protection from electrical shock and personal injury this meter provides to the operator. Perform only the maintenance task you are qualified to do.
Periodically clean your meter's case using a damp cloth. DO NOT use abrasive, flammable liquids, cleaning solvents or strong detergents as they may damage the finish, impair safety or affect the reliability of the structural components.
Remove screws from battery compartment cover on back of meter and remove cover. Replace batteries with fresh batteries paying attention to polarity position. Replace cover and screws.
Range | Resolution | Accuracy | Overload Protection |
400mV | 0.1mV | ±(0.5% + 4 dgts) | 1000V RMS |
4V | 1mV | ||
40V | 10mV | ||
400V | 100mV | ||
1000V | 1V | ±(0.8% + 10 dgts) |
Range | Resolution | Accuracy | Overload Protection |
400mV | 0.1mV | ±(2.0% +5 dgts) | 750V RMS |
4V | 1mV | ||
40V | 10mV | ||
400V | 100mV | ||
750V | 1V |
Range | Resolution | Accuracy | Overload Protection |
40A | 0.01A | ±(2.9% + 15 dgts) | 400A |
400A | 0.1A | ±(1.9% + 8 dgts) |
DL369: 400A range only
DL389: 45Hz to 400Hz True RMS (Crest factor<3:1)
Range | Resolution | Accuracy | Overload Protection |
400µA | 0.01µA | ±(1.2% + 3 dgts) | 2000µA/600V RMS |
2000µA | 0.1µA |
Range | Resolution | Accuracy | Overload Protection |
400µA | 0.01µA | ±(2.0% + 5 dgts) | 2000µA/600V RMS |
2000µA | 0.1µA | ±(1.5% + 5 dgts) |
DL389: 45Hz to 400Hz True RMS (Crest factor<3:1)
Range | Resolution | Accuracy | Overload Protection |
400Ω | 100mΩ | ±(1.0% + 4 dgts) | 600V RMS |
4kΩ | 1Ω | ||
40kΩ | 10Ω | ||
400kΩ | 100Ω | ||
4MΩ | 1kΩ | ||
40MΩ | 10kΩ | ±(2.0% + 4 dgts) |
Range | Resolution | Accuracy | Overload Protection |
40nF | 0.01nF | ±(3.5% + 6 dgts) | 600V RMS |
400nF | 0.1nF | ||
4µF | 0.001µF | ||
40µF | 0.01µF | ||
400µF | 0.1µF | ||
4000µF | 1µF |
DL379B: Capacitance range to 2000µF
Range | Open Circuit V | Test Current | Overload |
2.0V | <1.6V DC | 0.25mA | 600V RMS |
Range | Resolution | Accuracy | Overload Protection |
9.999Hz | 0.001Hz | ±(0.1% + 4 dgts) | 600V RMS |
99.99Hz | 0.01Hz | ||
999.9Hz | 0.1Hz | ||
9.999kHz | 1Hz | ||
99.99kHz | 10Hz | ||
199.9kHz | 100Hz |
Minimum Frequency: 0.5Hz, DC V offset should be zero
Sensitivity: >10% of each AC Volt range except 4V (>20%) range only
Range | Resolution | Accuracy | Overload Protection |
-22˚ ~ 14˚F | 0.1˚F | ±(1.0% +5.4˚F) | 30V RMS |
15˚ ~ 752˚F | ±(1.0% + 3.6˚F) | ||
(-30˚ ~ -10˚C) | (0.1˚C) | (1.0% +3.0˚C) | |
(-9˚ ~ 400˚C) | (±1.0% + 2.0˚C) |
Range | Accuracy | Overload Protection |
01. to 99.9% | ±(0.2% per kHz +0.1%) + 5 counts | 600V RMS |
0.5Hz to 100kHz (pulse width >2µsec)
The DL369/DL379B/DL389 is warranted to be free from defects in materials and workmanship for a period of three years from the date of purchase. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi's option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance. Batteries and consequential damage resulting from failed batteries are not covered by warranty.
Any implied warranties, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss.
A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered. Instruments out of warranty will be repaired (when repairable) for a service charge.
This warranty gives you specific legal rights. You may also have other rights which vary from state to state.