BEFORE TESTING

CHECKLIST
o Clean particle filter
o Water trap and probe line are
o All hose and thermocouple
o Water trap is fitted correctly
o Power on and zero
o Flue gas probe is sampling

COMBUSTION ANALYSIS

Insert flue probe into combustion
input. Insert K-type temperature
probe into input T1.
NOTE: If there is no hole, check with service and installation instructions of appliance for best testing
method or check with your local authority on best practice
O2/Eff displays CO2, O2,
Efficiency, Flue temperature,
Inlet temperature.
Find Quality Products Online at:
free of condensate
connections are properly
secured
Remove the rubber boot and red
drain plug. Allow the water to
drain out. Re-insert the rubber
ambient FRESH air
plug and replace boot cover.
T2
T1
P2
P1
Insert flue probe into stack.
Adjust cone so the probe tip is
approximately at the center of
the stack.
Rotate the function rotary dial
to CO/CO2 for Test Screen 2
CO/CO2
O2/Eff
Temp
parameters.
Exch Test
Aux
Status
v3
C163
GlobalTestSupply
www.
Remove protective rubber boot.
Remove water trap from analyzer.
Remove particle filter from water
trap and replace with new one.
Reattach trap and boot.
Rotate function rotary dial
to O2/Eff for Test Screen 1
parameters.
CO/CO2
O2/Eff
Temp
Exch Test
Aux
Status
v3
C163
CO/CO2 displays CO (ppm), CO2,
Losses calculated, Excess air %.
Make any adjustments needed
for proper Combustion and wait
for analyzer to display change in
readings. (Repeat as necessary).
CO/CO2
O2/Eff
Temp
Aux
Exch Test
Status
v3
C163
CO/CO2
O2/Eff
Temp
Exch Test
Aux
Status
v3
C163