Section 9
Troubleshooting
Miscellaneous Messages
AUDIO OFF DISABLED
BATTERY VERY LOW
PLUG IN AC POWER
EVENT MARKED
MONITOR PERFORM-
ING SELF TEST.
Parameters Reset
To Factory Default
DSP SERIAL TIMEOUT
DSP NOT RESPONDING
DSP ERROR
44
Model 2001 Service Manual
Displayed if user tries to enable Audio Off mode (by pressing and
holding the AUDIO key) while the "Allow Audio Off" portion of the
Options Menu is set to "No".
1. Monitor is running on battery power and the battery power has
been depleted. Connect line cord to AC Mains power source and
set the rear panel switch to "|"
2. Monitor's rear panel fuse has blown, monitor switched over to
battery power and has depleted battery life. Contact qualified
service personnel.
An event was successfully entered into trend memory.
Monitor is performing its power up system diagnostic tests.
Displayed when monitor is turned on while pressing the
ALERT RESET
during power on. Monitor now using factory default settings.
The main microprocessor has lost communication with the Digital
or
Signal Processor. This message will be displayed for 10 seconds
then the monitor will reset. The error should be recorded and
reported to service if the message is persistent.
System error has been detected on the Main Board. This message
will be displayed for 10 seconds then the monitor will reset. The
error should be recorded and reported to service if the message is
persistent.
key, or if an error found in battery-backed RAM
Rev. 01
Need help?
Do you have a question about the MARS PO2 TECH 2001 and is the answer not in the manual?