15. Maintenance and Inspection

Immediately remove faulty/defective medical beds from operation and secure them against
unauthorised use (disconnect the mains plug).
Immediately show any defects to the responsible personnel.
It is imperative that you always rectify any faults which could influence the function and
safety of the nursing bed immediately.
Observe the safety instructions contained in Section 5.
15.1 Maintenance – daily
The daily maintenance work can be performed by the nursing personnel or carer:
Lower both bed end actuator motors until they switch off automatically.
Check the side rails (option) for easy movement and damage.
Check the correct function of the locking device on the side rails (option).
All moving parts on this bed are fitted with maintenance-free self-lubricating bearings and
should not be oiled or greased.
Check to ensure that all cables are located safely, not dangling down, and not at risk of
being caught in the bed mechanism, by a mobile hoist or any other equipment that may
be used under the bed.
Whenever the bed is moved
Check the mains cable and plug for damage and kinks.
Route the mains cable in such a way that it cannot be pinched or rolled over.
Check to ensure that all cables are located safely, not dangling down, and not at risk of
being caught in the bed mechanism, by a mobile hoist or any other equipment that may
be used under the bed.